S6 JAGUAR

Technical Details

S6 JAGUAR EasyLoad — S6 JAGUAR EasyLoad ONLINE — S6 JAGUAR 15 Position XY changer


SpecificationBenefit
HighSense X-ray Power400 W
50 kV max.
17 mA max.
Analytical Performance (better precision and detection
limits), optimal heavy element analysis and best light
element analysis with low kV, high mA at full power
Sample SizesMask diameter: 34 mm standard
Optional additional sample masks:
28 mm, 23 mm, 18 mm
Maximum intensity and analytical flexibility for
different sample diameters
Primary beam filterAutomatic 5-position beam
filter changer (optional)
SampleCare™ tube protection
Better detection limits due to enhanced
peak-to-background ratio
Tube shield at sample loading
Analyzer crystalsXS-55* for F –Mg
PET* for Al – Cl
LiF200* for K – U
Ge*, XS-Ge-C*, XS-400*, LiF220*
Analytical flexibility for the entire element range


Enhanced performance
DetectorsProportional counter*
HighSense XE* for Ti – Am
Optimal detection
Linear range: 2 Mcps
Vacuum pumpIncluded, 265 mm x 180 mm x 489 mmLow cost of operation
Helium purge(optional)Liquid analysis
TouchControl™Intuitive user interface (optional)
Free language selection
Failsafe operation and reduces operator training
Temperature control unitIncluded, no additional chiller
466 mm x 190 mm x 630 mm
High analytical precision, better temperature stability
Power Supply100 – 240 V, 50/60 Hz, max. 1 kVAStandard wall plug
SafetyDIN EN ISO 9001:2008 CE certified
German type approval, radiation < 1 μSv/h

S6 JAGUAR
Manual

 

S6 JAGUAR
EasyLoad

S6 JAGUAR
EasyLoad ONLINE
S6 JAGUAR
15 Position XY changer
Manual sample loading

24-position XY-autochanger
for 51.5 EasyLoad rings

22-position XY-autochanger
for 51.5 EasyLoad rings
Autochanger for two-part
sample cups
51 x 67 x 74 cm
(height x width x depth);
132 kg
80 x 67 x 74 cm
(height x width x depth);
147 kg
80 x 67 x 74 cm
(height x width x depth);
147 kg
80 x 67 x 74 cm
(height x width x depth);
147 kg

* optional packages