banner-recirculation_compressor.jpg

Cryogen-Free Helium Recirculation System

New development to reduce consumable cost for ER 4112HV-CFxx

  • Additional compressor recirculates exchange gas
  • Adds 3 kW to power/water requirement
  • Purifier to remove Oxygen / Nitrogen
recirculation_compressor.jpg
Recirculation Compressor

Cryogen-Free EPR VT Systems

  • ER 4112HV-CF42 In-Cavity Cryogen-Free System Base temperature < 4 K to > 150 K
  • ER 4112HV-CF100 In-Cavity Cryogen-Free System Base temperature < 10 K to > 150 K
  • ER 4118HV-CF58 FlexLine Cryogen-Free System Base temperature < 5.8 K to > 150 K
  • ER 4118HV-CF100 FlexLine Cryogen-Free System Base temperature < 10 K to > 150 K

System Requirements

  • Systems use 99.999% purity helium gas for thermal transfer
  • Additional 7 kW Power and Cooling requirement to system specifications
  • Available in 208VAC 3-phase and 400VAC 3-phase power
ER 4118CF-O cryostat mounted with the EN 4118X-MD4 Pulse ENDOR Resonator
ER 4112HV-CF100 Installed in 10” Magnet

ER 4112HV-CF42 Cryogen-Free System

time_course_295k.PNG
In-Cavity Cooldown Profile
Related Information