banner_Biospec-152_11.png

General Specifications

 

Superconducting Magnet System

  • Actively-shielded superconducting wide-bore magnet
  • Nitrogen free (no cryogenic maintenance by customer)
  • Cryo-refrigerator (Minimal helium consumption, long hold-times, long maintenance intervals)
  • Reduced stray field
  • Screening against external field perturbations
  • Table attached to the magnet front side with integrated rail for animal handling system
  • Passive RT shim system
  • Vibration damping (absorber)
     

Electronics

  • Bruker Avance digital radio frequency system
  • The digital signal processing system operates at 1H resonance frequency of 500 MHz and utilizes digital filtering, digital quadrature demodulation as well as digital controls for RF amplitude stabilization for superior resolution and image quality
  • Receiver bandwidth: 5 MHz
  • Digital receive channels: up to 16 (optional)
  • 1H RF amplifier: 4000 W pulse power
  • HP workstation
  • High resolution flat screen monitor

BioSpec 152/11 USR/R

Superconducting Magnet System

  • Field strength: 15.2 T
  • Diameter of clear bore: 110 mm
  • Stray-field (5 Gauss): +/- 2.3 m axial,
    +/- 2.5 m radial
  • Length: 1.591 m
  • Diameter: 1.362 m

Gradient system B-GA 6S-100 with RT-shim and related power supplies

  • Actively shielded gradients
  • Integrated 2nd to 3rd shim set
  • Outer / inner diameter: 102 mm / 60 mm
  • Gradient amplitude: 1000 mT/m
  • Slew rate: 9000 T/m/s
  • Linearity 40 mm DSV: ± 5%
  • Water cooled

BioSpec 117/16 USR

Superconducting Magnet System

  • Field strength: 11.7 T
  • Diameter of clear bore: 160 mm
  • Stray-field (5 Gauss): +/- 2.8 m axial,
    +/- 1.7 m radial
  • Length: 1.46 m
  • Diameter: 1.652 m

Gradient system B-GA 9S with RT-shim and related power supplies

  • Actively shielded gradients
  • Integrated shim set up to 2nd order
  • Outer / inner diameter: 152 mm / 90 mm
  • Gradient amplitude: >750 mT/m
  • Slew rate: 6660 T/m/s
  • Linearity 60 mm DSV: < ± 5%
  • Water cooled

BioSpec 94/20 USR

Superconducting Magnet System

  • Field strength: 9.4 T
  • Diameter of clear bore: 200 mm
  • Stray-field (5 Gauss): +/- 3 m axial,
    +/- 2 m radial
  • Length: 1.49 m
  • Diameter: 1.652 m

Gradient system B-GA 12S2 with RT-shim and related power supplies

  • Actively shielded gradients
  • Integrated shim set up to 2nd order
  • Outer / inner diameter: 205 mm / 116 mm
  • Gradient amplitude: 440 mT/m
  • Max. slew rate: 3440 T/m/s
  • Linearity 80 / 50 mm DSV: < ± 4.5%,
    < ± 1.0%
  • Water cooled

BioSpec 47/40 USR

Superconducting Magnet System

  • Field strength: 4.7 T
  • Diameter of clear bore: > 400 mm
  • Stray-field (5 Gauss): +/- 3 m axial,
    +/- 2 m radial
  • Length: 1.49 m
  • Diameter: 1.652 m

BioSpec 117/11 USR/R

Superconducting Magnet System

  • Field strength: 11.7T
  • Diameter of clear bore: 110 mm
  • Stray-field (5 Gauss): +/- 1.8 m axial,
    +/- 1.25 m radial
  • Length: 1.3 m
  • Diameter: 1.242 m

Gradient system B-GA 6S-100 with RT-shim and related power supplies

  • Actively shielded gradients
  • Integrated 2nd to 3rd shim set
  • Outer / inner diameter: 102 mm / 60 mm
  • Gradient amplitude: 1000 mT/m
  • Slew rate: 9000 T/m/s
  • Linearity 40 mm DSV: ± 5%
  • Water cooled

BioSpec 94/30 USR

Superconducting Magnet System

  • Field strength: 9.4 T
  • Diameter of clear bore: 300 mm
  • Stray-field (5 Gauss): +/- 3.3 m axial,
    +/- 2.3 m radial
  • Length: 2.08 m
  • Diameter: 1.706 m

Gradient system B-GA 20S with RT-shim and related power supplies

  • Actively shielded gradients
  • Integrated shim set up to 2nd order
  • Outer / inner diameter: 301.5 mm / 200.5 mm
  • Gradient amplitude: 200 mT/m
  • Max slew rate: 640 T/m/s
  • Linearity 130 / 100 mm DSV: < ± 3%,
    < ± 2%
  • Water cooled

BioSpec 70/20 USR

Superconducting Magnet System

  • Field strength: 7 T
  • Diameter of clear bore: 200 mm
  • Stray-field (5 Gauss): +/- 1.5 m axial,
    +/- 1.5 m radial
  • Length: 1.312 m
  • Diameter: 1.12 m

Gradient system B-GA 12S2 with RT-shim and related power supplies

  • Actively shielded gradients
  • Integrated shim set up to 2nd order
  • Outer / inner diameter: 198 mm / 114 mm
  • Gradient amplitude: 440 mT/m
  • Slew rate: 3440 T/m/s
  • Linearity 80 / 50 mm DSV: < ± 4.5%,
    < ± 1.0%
  • Water cooled

BioSpec 70/30 USR

Superconducting Magnet System

  • Field strength: 7 T
  • Diameter of clear bore: > 300 mm
  • Stray-field (5 Gauss): +/- 3 m axial,
    +/- 2 m radial
  • Length: 1.49 m
  • Diameter: 1.652 m

Gradient system B-GA 20S with RT-shim and related power supplies

  • Actively shielded gradients
  • Integrated shim set up to 2nd order
  • Outer / inner diameter: 301.5 mm /
    200.5 mm
  • Gradient amplitude: 200 mT/m
  • Max. slew rate: 640 T/m/s
  • Linearity 130 / 100 mm DSV: < ± 3%,
    < ± 2%
  • Water cooled