G4 ICARUS Series 2

G4 ICARUS Series 2 - Technical Details


SpecificationBenefit
Samples
Sample Types

Inorganic, dry, any form or shape

Flexibility
Typical Samples
Ferrous and nonferrous metals and their alloys, ores, cement, limestone, glass, ceramics, soilsWide variety of sample materials
Sample carrier
Ceramic crucibles (standard form)Usage of industry standard consumables, lower cost, security of supply
Sample weight
0.1 – 1 g (typical 0.5 g)
Furnace
High frequency induction furnace, power adjustable with ZoneProtect™More efficient, faster combustion with maximum productivity

Detectors
Carbon
Sulfur

Dual-range NDIR detector with reference beam path
HighSense™ NDUV2) detector with reference beam path


Reliable, maintenace-free performance without drift
Unique wide range performance with unrivaled signal/noise without drift

Carrier Gas
Oxygen 99.95% purity
(99.995% required for trace carbon)
3.5 bar (50 psi) min. pressure
Standard gas purity sufficient
Compressed air
5 bar (72.5 psi) min. pressure, oil-, water- and particle-free
Speed
Analysis Time

40 s (nominal)

Fast, simultaneous analysis
Complete Cycle Time
< 1 min (typical)High productivity and throughput with waste disposal into crucible
Noise Emission
< 50 bdEnjoy the silence
Calibration
Single and multipoint calibration with solid standards
Reagents
  • Magnesium perchlorate
  • Sodium hydroxide on support
  • Platinized silica
Power Supply
230 VAC (± 10 %); L1 N; -> 50-60 Hz; 16 A; 2.7 kVA
208 VAC (± 10 %); L1 L2; -> 50-60 Hz; 16 A; 2.7 kVA
Industry standard power and current configurations

2) NDIR-based detector available on request