Systems | S8 TIGER 1 kW 1kW 50 kV max. 50 mA max. | S8 TIGER 3 kW 3 kW 60 kV max. 150 mA max. | S8 TIGER 4 kW 4 kW 60 kV max. 170 mA max. |
X-ray tubes | Rh, 75 μm Be Window | Rh, 75 μm Be Window Rh, 50 μm Be Window, SampleCare protective coating Rh, 28 μm Be Window, SampleCare protective coating Cr, max.3.3 kW | Rh, 75 μm Be Window Rh, 50 μm Be Window, SampleCare protective coating Rh, 28 μm Be Window, SampleCare protective coating Cr, max.3.3 kW |
Collimator | Automatic collimator changer (up to 4) | | |
Analyzer crystals | Automatic crystal changer (up to 8) Included: XS-55, PET, LiF (200) Optional: XS-B, XS-C, XS-N, XS-PET-C, XS-CEM, XS-Ge-C, LiF (220), LiF (420), ADP, Ge, TlAP, InSb, XS-N-HS, XS-100, XS-400, XS-InSb-C | | |
DynaMatch™ | Linear intensity range more than 13 million cps | | |
HighSense Detectors | Proportional flow counter and Scintillation Counter: Capacity more than 3 Mio cps, 2 Mio cps linear range | | |
XRF2 | High Resolution and high Intensity Element Mapping with 0.3 mm and 1.25 mm spot (FWHM) | | |
Automated Mask Changer | Automatic masks : closed (SampleCare;) 34 mm, 28 mm, 23 mm, 18 mm, 8 mm (BeamGuide), 5 mm (BeamGuide) | | |
TouchControl™ 1) | Integrated touchscreen for easy and intuitive operation | | |
SampleCare™ 1) | X-ray tube and goniometer protected by contamination shields Sample and spectrometer chamber separated by programmable vacuum seal | | |
EasyLoad™ 1) | Automatic sample recognition Portable sample trays | | |
Power requirements | 208 – 240 V (1P/3P) 50/60 Hz | 208 V, 60 Hz (1P/3P) 230 V, 50/60 Hz (3P) | |
Compressed air | Not required | | |
Detector gas | | P5 gas (5% methane, 95% argon) for flow counter P10 gas (10% methane, 90% argon) for flow counter | |
External cooling water | No cooling water | Cooling water Water consumption automatically regulated and minimized, short term interruptions are compensated | |
Dimensions (height x width x depth) | 135 cm x 84 cm x 90 cm; 53.1” x 33.1” x 35.4” | 135 cm x 84 cm x 104 cm; 53.1” x 33.1” x 41” | |
| Touchscreen: Allows additional width of 49 cm (19.3’’) | | |
| 446 kg | 476 kg | | |
Quality & safety | DIN EN ISO 9001:2008 CE certified Fully radiation protected system; radiation < 1 μSv/h (BfS 09/07 V RöV) | | |