S2 POLAR - Technical Details

S2 POLAR Refinery Analyzer


SpecificationBenefit
ApplicationsElemental analysis in refineries e.g. sulfur analysis of
  • gasoline, diesel
  • biodiesel
  • kerosene, jet fuel, heating oil
  • naphtha, residual oil to crude
Further elements on request (e. g. P, Cl, Fe, Ni, V)
All refinery applications on one unit
NormsASTM D7220-17: Sulfur in automotive, heating, and jet fuels from 3 ppm to 942 ppm
ASTM D4294-16: Sulfur in petroleum and petroleum products from 16 ppm to 5 %
Ready-to-analyze solutions* for ASTM D7220 and D4294 including blanks, set of standards, QC and DC samples
Fulfills ISO 13032, 20847, 8754, IP 336, 496, 532,
and JIS K 2541-4

Norm-compliant sulfur analysis, internationally accepted, including ultra-low sulfur (ULS) applications

 

Dedicated, optional pre-installed push button methods to fit for purpose

Detection Limit (LLD)0.7 ppm S at 300 s measurement timePrecise and accurate S measurements, including ultra-low sulfur (ULS). Enables thesh-hold relevant process control at all steps in refineries
Measurement Range3 ppm to 5 % S, combined in one calibration with automated line switch
Higher concentrations on request
One calibration with wide concentration range

S2 POLAR Oil Analyzer


SpecificationBenefit
ApplicationsElemental analysis of additives in oils and polymersElemental analysis optimized for petrochemical products
NormsASTM D6481-14: P, S, Ca, Zn,
ASTM D7751-16: Mg, P, S, Cl, Ca, Zn, Mo
Ready-to-analyze solutions* for ASTM D6481 and D7751 including blanks,
set of standards, QC and DC samples
Further elements and norms on request

Norm-compliant analysis, internationally accepted

Dedicated, optional pre-installed push button methods to fit for purpose

S2 POLAR Refinery & Oil Analyzer


SpecificationBenefit
Atmosphere Modes
Helium mode
Vacuum mode
Optimal light element analysis of liquids
Low cost of operation
Sample Preparation
Liquid cups, SampleCare cups, Prolene and Mylar® foils, pipettes, balanceAccessories ensure high throughput of liquid samples. Low-cost per sample due to standardized liquid cups
Further Options

Emergency Machine Off (EMO)
Uninterruptible Power Supply (UPS)
Sample rotation

 

Compliant with safety requirements
Enables removing of liquid samples
Enhanced precision for inhomogenous samples, such as polymer pucks
X-ray Tube
50 W, high-power X-ray tube, max. voltage 50 kV, with polarizing HighSense™ beam path

Optionally: 30 kV max.
Max. power for short measurement times and high sample throughput, beam path optimized for petrochemical materials
Simplify regulatory efforts (e.g. Austria, France, Italy, Taiwan)
DetectorHighSense™ ULS Silicon Drift DetectorHighest count rates for fast analysis, low LLD
TouchControl™Integrated 12.1‘‘ TFT touchscreen, multilingual user interface: English, German, French, Spanish, Portuguese, Italian, Russian, Chinese, JapaneseIslandMode™ without external PC
Intuitive and easy-to-use, in your own language
ConnectivityEthernet port RJ45, 3x USB ports for mouse, keyboard, and printer; HDMI/VGA ports for external display, remote access via TCP/IPIslandMode™ but not isolated, various options for printing and network data transfer, even fully remotely
Power Supply100-240 V, 50/60 Hz, max. 600 VAStandard wall plug
Dimensions; width x depth x height, weight46.6 x 74.5 x 37.0 cm, 55 kg
18.3’’ x 29.3’’ x 14.6’’, 121 lbs
Small and compact for installations with limited space, e.g. for on-site process control in refineries
Safety

DIN EN ISO 9001:2008, 2006/42/EC (CE-certified Machinery directive), 2014/35/EC (Electrical equipment), 2014/30/EC (Electromagnetic Compatibility),
German Type Approval and Vollschutz according to BfS RöV,
Fully radiation-protected system; radiation <1 μSv/h (H*), Compliant to ICRP, IAEA, EURATOM

* Optional packages
Mylar® is a trademark of DuPont Nemours Inc.
HighSense™, IslandMode™, SampleCare™, TouchControl™ are trademarks of Bruker AXS.