Fusing Magic Angle Spinning (MAS) technology with the MR Microscopy gradient technology enables to acquire images and localized spectra of semi-solid objects due to the combination of reduced signal linewidth with a high and very linear gradient field. Rotors with a 3.2 or 4 mm diameter are offered, further on request. Besides imaging applications very slow diffusion processes can be investigated applying the strong gradient fields.
It is combined with Bruker’s Micro2.5 WB gradient system, which is the MR Microscopy gradient workhorse. Its gradient strength is perfect for most applications. Utilizing the Micro2.5 gradient the maximum gradient strength of 2.5 T/m along the magic angle is reached if the x, y, z gradients are switched simultaneously.
The MR Microscopy MAS Probe is a very dedicated and customized device. Please contact the MR Microscopy Product Management before placing an order.
The MR Microcopy Micro2.5 Magic Angle Spinning probe offers unique capabilties for imaging and diffusion experiments of semi-solid objects.
Micro2.5 Magic Angle Spinning | |
Bore | WB / SWB |
1H frequency range | 300 – 800 MHz |
rf channels | Two independent |
Requirements | Microscopy Accessory |
Gradient | Micro2.5 |
Directions | xyz |
Gradient strength per direction | 1.5 T/m |
ID/OD | 40/72 mm |
Rise Time (5-95%), 0-60A,120V | < 100 µs |
Cooling | Water |
Maximum current tested | 60 A |
Removable | Yes |
Bruker’s commitment to provide customers with unparalleled help throughout the buying cycle, from initial inquiry to evaluation, installation, and the lifetime of the instrument is now characterized by the LabScape service concept.
LabScape Maintenance Agreements, On-Site On-Demand and Enhance Your Lab are designed to offer a new approach to maintenance and service for the modern laboratory