New SA Instruments Animal Monitoring Software Version

SA Instruments has recently released a new PC-SAM software version (12.00) for Models 1030, 1025T and 1025 and recommends users to update to this latest version. 
PC-SAM 12.00 resolves potential issues with moving windows.

For ERT 1030 Modules with firmware version 38, previous software versions display systematically too high (Version < 10.03) or too low (10.3 ≤ Version < 11.00) temperature readings. The offset is in the range of 0.7°C at 37°C.

PC-SAM 12.00 is available from the vendor website for download on https://i4sa.com/ (external link).